Skip to main content

Table 6 Breakdown of the uncertainty budget

From: Development of an ID-LC–MS/MS method using targeted proteomics for quantifying cardiac troponin I in human serum

 

TLLQIAK (μg/L)

NITEIADLTQK (μg/L)

2

5

12

2

5

12

Mean

2.1

4.9

11.4

2.1

5.0

12.7

Uncertainty components

sr

0.1

0.2

0.3

0.2

0.2

0.5

sbetween

0.1

0.3

0.2

0.2

0.2

0.3

ucal

0.2

0.2

0.2

0.2

0.2

0.2

Balance

0.0

0.0

0.0

0.0

0.0

0.0

k

2

2

2

2

2

2

Uncertainty, uc

0.3

0.5

0.5

0.3

0.4

0.6

Expanded uncertainty, U

0.6

0.9

0.9

0.7

0.7

1.2